HEATER COOLER 3T (230 VOLT)
Report
- Report Number
- 9611109-2024-00026
- Event Type
- Death
- Date Received
- January 11, 2024
- Date of Event
- August 22, 2012
- Report Date
- June 27, 2024
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
H11: LEGAL LAWSUIT HAS BEEN ISSUED AND NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE. A DEVICE HISTORY RECORD (DHR) REVIEW COULD NOT BE PERFORMED SINCE POSSIBLE SERIAL NUMBERS INVOLVED WERE NOT PROVIDED. INVOLVED DEVICE SERIAL NUMBER REMAINED UNKNOWN BUT IT WAS NOT EQUIPPED WITH VACUUM AND SEALING KIT SINCE UPGRADE ACTIVITY STARTED IN 2017 AND SURGERY WAS PERFORMED IN 2012. NO FURTHER INVESTIGATION IS POSSIBLE. SOURCE OF PATIENT CONTAMINATION REMAINS UNKNOWN AND A DIRECT RELATIONSHIP BETWEEN THE REPORTED ADVERSE EVENT AND THE DEVICE COULD NOT BE ESTABLISHED.
G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). H.4. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. H.9. LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN ITALY. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. THIS REPORT WAS DUE ON DECEMBER 02, 2023; HOWEVER, DUE TO A NETWORK DISRUPTION AT LIVANOVA, THE ABILITY TO SUBMIT MDRS WAS LOST ON NOVEMBER 19, 2023, BEFORE THIS REPORT WAS READY TO SUBMIT. THE REPORT WAS PREPARED AND SUBMITTED FOLLOWING RESTORATION OF THE LIVANOVA SYSTEMS.
SEE INITIAL REPORT.
LIVANOVA DEUTSCHLAND RECEIVED, THROUGH A LAWSUIT, A REPORT THAT A MALE PATIENT UNDERGONE SURGERY ON (B)(6) 2012 WAS FOUND INFECTED BY MYCOBACTERIUM CHIMAERA AND DIED ON (B)(6) 2013. HEATER-COOLER SYSTEM 3T WAS USED DURING SURGERY. BASED ON CURRENT STATUS OF THE INVESTIGATION THE ALLEGED DEVICE ISSUE WAS NOT CONFIRMED YET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 667912 | HEATER COOLER 3T (230 VOLT) | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Death |