IMP,TSV,3.7,11.5,MTX,MG
Report
- Report Number
- 0002023141-2024-00032
- Event Type
- Injury
- Date Received
- January 10, 2024
- Date of Event
- November 20, 2023
- Report Date
- June 13, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019935
- PMA / PMN Number
- K101977
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4109, 4111 AND 3331. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 3252. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4307. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED IMPLANT FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED. IMPLANT FRACTURE IDENTIFIED AT THE COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1242846. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1242846 FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD ¿ PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. THE IMPLANT¿S COLLAR WAS FRACTURED, AND THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE PATIENT CAME IN FOR A BROKEN IMPLANT COLLAR/SCREW. THE IMPLANT WAS REMOVED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645008 | IMP,TSV,3.7,11.5,MTX,MG | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1242846 | 00889024019935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Female | Required Intervention |