FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3ML LL BNS

MDR report key: 18461465 · Received January 7, 2024

Report

Report Number
1213809-2023-01518
Event Type
Malfunction
Date Received
January 7, 2024
Date of Event
December 11, 2023
Report Date
May 9, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
50382903010733
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

TWO PHOTOS OF A 3ML LUER-LOK SYRINGE WERE RECEIVED AND EVALUATED. BOTH PHOTOS WITH ONE A CLOSE-UP SHOWS A LOOSE SYRINGE FILLED WITH AN UNKNOWN FLUID WITH AN UNKNOWN BROWN PARTICULATE LARGER THAN LEVEL 4 IN THE NON-FLUID PATH. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER NON-FLUID PATH DEFECT IS ASSOCIATED WITH THE BULK HANDLING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 3143639 SHOWING NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SYRINGE 3ML LL BNS HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN DRAWING THIS SYRINGE, I NOTICED WHAT APPEARS TO BE PIECE OF CARDBOARD/DIRT." RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. COMPLAINT NUMBER (B)(4). PRODUCT CODE BX301073. LOT NUMBER 2297482, 3143639. PRODUCT DESCRIPTION UNIVERSAL ANES.TRAY MRX. COMPLAINT DESCRIPTION WHEN DRAWING THIS SYRINGE, I NOTICED WHAT APPEARS TO BE PIECE OF CARDBOARD/DIRT. SAMPLE AVAILABILITY NO. PATIENT INJURY NO. SHOULD YOU HAVE ANY MORE QUESTIONS PLEASE LET ME KNOW.

Description of Event or Problem · 0

MATERIAL#: 301073, LOT#: 2297482,3143639. IT WAS REPORTED BY THE CUSTOMER REPORTED THAT WHEN DRAWING THIS SYRINGE, I NOTICED WHAT APPEARS TO BE PIECE OF CARDBOARD/DIRT. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. COMPLAINT NUMBER (B)(4). PRODUCT CODE BX301073. LOT NUMBER 2297482, 3143639. PRODUCT DESCRIPTION UNIVERSAL ANES TRAY MRX. COMPLAINT DESCRIPTION WHEN DRAWING THIS SYRINGE, I NOTICED WHAT APPEARS TO BE PIECE OF CARDBOARD/DIRT. SAMPLE AVAILABILITY NO. PATIENT INJURY NO. SHOULD YOU HAVE ANY MORE QUESTIONS PLEASE LET ME KNOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
568360 BD SYRINGE 3ML LL BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 2297482 50382903010733

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown