VP 2000 PROCESSOR
Report
- Report Number
- 3005248192-2010-00008
- Event Type
- Malfunction
- Date Received
- September 27, 2010
- Date of Event
- August 9, 2010
- Report Date
- August 9, 2010
- Manufacturer
- ABBOTT MOLECULAR, INC
- Product Code
- KEY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
ABBOTT MOLECULAR INC. IS AWARE OF THE PROBLEM WITH THE VP 2000 PROCESSOR BASINS. AN URGENT PRODUCT RECALL (3005248192-06/18/2010-001-C) WAS DISTRIBUTED VIA (B)(4) TO (B)(6) CUSTOMERS BASED ON CUSTOMER RECEIPT OF VP 2000 PROCESSOR AND BASIN SPARE PART SINCE (B)(4) 2008. THE FOLLOWING MDRS WERE ALSO FILED: MDR 3005248192-2010-0002, FILED ON (B)(4). MDR 3005248192-2010-0005, FILED ON (B)(4). MDR 3005248192-2010-0006, FILED ON (B)(4). MDR 3005248192-2010-0007, FILED ON (B)(4). THREE POTENTIAL HAZARDS WERE IDENTIFIED: DELAY IN RESULTS. THERE IS NO HARM ASSOCIATED WITH THIS HAZARD. CHEMICAL EXPOSURE. THERE IS NO HARM ASSOCIATED WITH THIS HAZARD AS THIS ISSUE DOES NOT INCREASE THE ALREADY EXPECTED EXPOSURE TO THESE REAGENTS. THEREFORE, THERE IS NO INTRODUCED OPERATOR HARM. SLIP HAZARD. EVALUATION DETERMINED THAT THE OVERALL HEALTH RISK WAS LOW, BASED ON THE FOLLOWING: THE OVERALL ESTIMATE FOR A LEAKING BASIN WAS (B)(4) (BASED ON BASINS KNOWN TO BE RELATED TO THIS ISSUE). (B)(4). WHEN A LOADED BASIN IS EITHER USED IN THE VP2000 OR STORED ON THE LAB BENCH, IN THE TYPICAL USAGE TIME < 8 HOURS), THE LEAKED LIQUID WILL BE LESS THAN 50 ML. SLIP HAZARD WILL BE AVOIDED BY THE OPERATOR'S OBSERVANCE OF THE LIQUID ON THE FLOOR AND BY THE SMALL VOLUME OF LIQUID. IF A LEAK WERE TO OCCUR, IT WOULD OCCUR IN AN AREA THAT A LAB OPERATOR WOULD NOT NORMALLY WALK (IN THE VP2000) OR DURING REAGENT ADDITION (LAB BENCH). IN ADDITION, THE LEAK WOULD BE VISUALLY OBVIOUS TO THE OPERATOR. IF THE OPERATOR DID STEP ON THE LEAK, IT WOULD NOT NECESSARILY LEAD TO A FALL THAT WOULD LEAD TO A SERIOUS INJURY.
THE ABBOTT VP 2000 PROCESSOR IS A DEVICE DESIGNED TO AUTOMATE AND STANDARDIZE SLIDE SPECIMEN PROCESSING AND ROUTINE SLIDE STAINING FOR THE LABORATORY. CUSTOMER REPORTED HAVING A LEAKING BASIN AFTER RECEIVING THE PRODUCT RECALL NOTIFICATION FOR POTENTIAL VP 2000 PROCESSOR LEAKING BASINS. THE NOTIFICATION ASKED FROM CUSTOMERS TO TAKE THE FOLLOWING ACTION: "IF YOU NOTICE LEAKING OR WARPED BASINS OR AN INABILITY TO PLACE THE BASINS INTO THE VP 2000 PROCESSOR, PLEASE CALL YOUR LOCAL ABBOTT MOLECULAR REPRESENTATIVE FOR COORDINATION OF BASIN REPLACEMENT. AN ABBOTT MOLECULAR CUSTOMER SERVICE REPRESENTATIVE OR FIELD SERVICE REPRESENTATIVE WILL BE CALLING YOU TO COORDINATE REPLACEMENT OF YOUR VP 2000 HEATED REAGENT BASINS BEFORE DECEMBER 31, 2010. PLEASE NOTE THAT ADDITIONAL MEASURES HAVE BEEN TAKEN TO ENSURE CURRENTLY AVAILABLE INVENTORY DOES NOT EXHIBIT THIS ISSUE." ABBOTT MOLECULAR DISTRIBUTED AN URGENT PRODUCT RECALL (3005248192-06/18/2010-001-C) VIA (B)(4) TO (B)(6) CUSTOMERS BASED ON CUSTOMER RECEIPT OF VP 2000 PROCESSOR AND BASIN SPARE PART SINCE (B)(6) 2008. THE FOLLOWING MDRS WERE ALSO FILED: MDR 3005248192-2010-0002, FILED ON (B)(4). MDR 3005248192-2010-0005, FILED ON (B)(4). MDR 3005248192-2010-0006, (B)(4). MDR 3005248192-2010-0007, FILED ON (B)(4). NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VP 2000 PROCESSOR | STAINER, TISSUE, AUTOMATED | KEY | ABBOTT MOLECULAR, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |