RESOLUTION HEMOSTASIS CLIPPING DEVICE
Report
- Report Number
- 3005099803-2010-03934
- Event Type
- Malfunction
- Date Received
- September 27, 2010
- Date of Event
- September 1, 2010
- Report Date
- September 6, 2010
- Manufacturer
- BOSTON SCIENTIFIC - MARLBOROUGH
- Product Code
- MND
- PMA / PMN Number
- K040148
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE PRONGS OF THE CLIP COULD NOT BE ACTUATED AND THERE WAS A KINK IN THE CONTROL WIRE NEAR THE HANDLE. FUNCTIONALLY, THE CLIP ASSEMBLY WAS THEN DEPLOYED, WITH TWO DISTINCTIVE CLICKS BEING HEARD, AS DESIGNED. THE ROOT CAUSE COULD NOT BE DETERMINED HOWEVER, THE MOST PROBABLE ROOT CAUSE HAS BEEN DETERMINED TO BE OPERATIONAL CONTEXT AS EVIDENT BY THE KINK IN THE CONTROL WIRE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT.
(B)(4). THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS HOWEVER, AN EVALUATION HAS NOT BEEN PERFORMED AS OF YET. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USED DURING A PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THEY DEPLOYED THE CLIP HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER. THE CASE WAS COMPLETED WITH ANOTHER RESOLUTION CLIP DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY. ATTEMPTS TO OBTAIN MORE COMPLETE INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USED DURING A PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THEY DEPLOYED THE CLIP HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER. THE CASE WAS COMPLETED WITH ANOTHER RESOLUTION CLIP DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY. ATTEMPTS TO OBTAIN MORE COMPLETE INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESOLUTION HEMOSTASIS CLIPPING DEVICE | LIGATOR, ESOPHAGEAL | MND | BOSTON SCIENTIFIC - MARLBOROUGH | M00522610 | 10060205C2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |