FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 1844771 · Received September 23, 2010

Report

Report Number
1720753-2010-03182
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
September 14, 2010
Report Date
September 23, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR IRIS POTENTIOMETER AND POTATIONAL BRAKE WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9600 SYSTEM HAD A COLLIMATOR IRIS POTENTIOMETER ERROR AT BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1