FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM

MDR report key: 18445866 · Received January 4, 2024

Report

Report Number
3014704491-2023-00874
Event Type
Malfunction
Date Received
January 4, 2024
Date of Event
November 4, 2023
Report Date
February 29, 2024
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903830336
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW(LOT#3080067): 1)THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN MAY 2023, AND PACKAGED AT R240 PACKAGE LINE IN MAY 2023. WORK ORDER QUANTITY WAS (B)(4). 2)REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3)REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE CATHETER OF THE RETAINED SAMPLE OF THIS BATCH OF IS EXAMINED UNDER MICROSCOPE, AND NO ABNORMALITY IS FOUND, (B)(4). 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS. AS THE DAMAGE STATE OF THE CATHETER IS NOT IDENTIFIABLE AND THE USAGE OF THE COMPLAINED SAMPLE IS UNKNOWN, THE ROOT CAUSE OF THE LEAKAGE CANNOT BE DETERMINED. H3 OTHER TEXT : SEE NARRATIVE.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM LEAKEDAFTER SUCCESSFULLY PUNCTURING A PATIENT, A LEAK IS FOUND IN THE INDWELLING NEEDLE HOSE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
559481 BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 3080067 00382903830336

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown