HIRES¿ ULTRA 3D IMPLANT
Report
- Report Number
- 3006556115-2023-02312
- Event Type
- Malfunction
- Date Received
- January 4, 2024
- Date of Event
- February 21, 2024
- Report Date
- December 18, 2023
- Manufacturer
- ADVANCED BIONICS, LLC
- Product Code
- MCM
- UDI-DI
- 07630016872037
- PMA / PMN Number
- P960058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
ADVANCED BIONICS CONSIDERS THE INVESTIGATION INTO THIS REPORTABLE EVENT AS CLOSED. ADVANCED BIONICS RECEIVED PERMISSION ON BEHALF OF THE RECIPIENT TO PROCEED WITH FAILURE ANALYSIS ON 04/11/2024. THE EXTERNAL VISUAL INSPECTION REVEALED THE ELECTRODE WAS SEVERED AND SILICONE DAMAGE WAS OBSERVED ON THE TOP COVER OF THE DEVICE. THESE ARE BELIEVED TO HAVE OCCURRED DURING REVISION SURGERY. THE DEVICE PASSED THE PHOTOGRAPHIC IMAGING INSPECTION. SYSTEM LOCK WAS VERIFIED. THE ELECTRODE CONDITION PREVENTED AN ELECTRICAL TEST FROM BEING PERFORMED. THE DEVICE PASSED SOME OF THE ELECTRICAL TESTS PERFORMED. THE FAILURE OF THIS DEVICE IS ATTRIBUTED TO AN ELECTRODE SHORT IN THE ELECTRODE POCKET. A CORRECTIVE ACTION WAS IMPLEMENTED. THIS VERSION OF THE ULTRA IS NO LONGER DISTRIBUTED. THIS IS THE FINAL REPORT. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
ADDITIONAL INFORMATION: SECTIONS B.3, D.6B THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. THE EXPLANTED DEVICE WAS RECEIVED BY ADVANCED BIONICS (B)(6) 2024. ADVANCED BIONICS IS CURRENTLY ATTEMPTING TO OBTAIN CONSENT FROM THE RECIPIENT. THE DEVICE REMAINS INTACT IN A LOCKED VAULT AT AB, LLC UNTIL CONSENT IS OBTAINED. DISCLAIMER: ADVANCED BIONICS DOES NOT INTEND THAT THIS REPORT BE ANY ADMISSION OF LIABILITY, FAULT OR PRODUCT DEFECT.
THE RECIPIENT IS REPORTEDLY EXPERIENCING DECREASED PERFORMANCE. A REVIEW OF THE RECIPIENT¿S TEST DATA INDICATED IMPEDANCE ISSUES. REVISION SURGERY WILL BE SCHEDULED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1501663 | HIRES¿ ULTRA 3D IMPLANT | COCHLEAR IMPLANT | MCM | ADVANCED BIONICS, LLC | CI-1601-05 | NA | 07630016872037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Female |