ANALYTICAL P MODULE
Report
- Report Number
- 1823260-2010-05695
- Event Type
- Malfunction
- Date Received
- September 24, 2010
- Date of Event
- September 10, 2010
- Report Date
- September 24, 2010
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K953239
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE USER RECEIVED QUESTIONABLE RESULTS FOR FERRITIN ON THE ANALYTICAL P MODULE FOR TWO PATIENT SAMPLES. PATIENT SAMPLE 1, THE INITIAL FERRITIN RESULT GAVE 12 PG/ML. THIS RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE SAMPLE WAS REPEATED GIVING 23 PG/ML. PATIENT SAMPLE 2, THE INITIAL FERRITIN RESULT GAVE 21 PG/ML. THIS RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE SAMPLE WAS REPEATED GIVING 39 PG/ML. NO ADVERSE EVENT HAS BEEN ALLEGED REGARDING THESE DISCREPANCIES. THE FERRITIN REAGENT LOT NUMBER WAS 62168101 ALTHOUGH THE FIELD APPLICATION SPECIALIST (FAS) COULD NOT DETERMINE A CAUSE, IT WAS NOTED MANY BUBBLES WERE FOUND IN THE FERRITIN R1 REAGENT PACK WHICH COULD INDICATE A REAGENT HANDLING OR SHIPMENT ISSUE AS THE CAUSE. THE FAS RAN PERFORMANCE TESTING WITH ALL RESULTS WITHIN SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANALYTICAL P MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |