GREENLIGHT ADDSTAT
Report
- Report Number
- 2937094-2010-00010
- Event Type
- Malfunction
- Date Received
- September 13, 2010
- Date of Event
- January 8, 2010
- Report Date
- January 11, 2010
- Manufacturer
- AMS INNOVATION CENTER-SILICON VALLEY
- Product Code
- GEX
- PMA / PMN Number
- K062719
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
A FAILURE ANALYSIS REPORT WAS GENERATED BY AN AMERICAN MEDICAL SYSTEMS QUALITY ENGINEER. THE EXAMINATION DISCLOSED THAT THE FIBER CAP REGION BURNED AND/OR MELTED, THAT THE CAP REMAINED INTACT AND ATTACHED AND THAT THE JACKET AND/OR SHRINK TUBE MAY HAVE BEEN MELTED AND/OR BURNT. THE EXAMINATION ALSO DISCLOSED THAT THE CAP GENERALLY EXHIBITED SOME OR ALL OF CHAR, DEVITRIFICATION, CRATER AND MELTED GLASS. THE DEVICE FAILURE WAS ASSOCIATED WITH A LACK OF COOLING. THE 2090 FIBER ENDURES A HIGHER POWER, HAS A REFLECTIVE CAP AREA, AND MAY BE SUBJECTED TO MORE HEAT, ESPECIALLY IF COOLING IS BLOCKED BY TISSUE CONTACT. TISSUE CONTACT CREATES CHAR WHICH FURTHER INSULATES THE CAP FROM COOLING AND INCREASES HEAT BY ABSORBING ENERGY. THE HEAT CONTRIBUTES TO DEVITRIFICATION AND ASSOCIATED WEAKENING OF THE GLASS, MAKING IT SUBJECT TO BREAKAGE FROM MELTING OR MECHANICAL OR THERMAL STRESS. GLUE BURNING INCREASES PRESSURE IN THE CAP. ENERGY REFLECTED BACK AT THE FIBER CAP FROM THE SCOPE, SEED, STONE OR OTHERWISE MAY HAVE CONTRIBUTED TO AND/OR CREATED ANY MELTING IN THE CRATER AREA OF THE FIBER CAP. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION OF THE FIBER. NO PT INJURY WAS REPORTED. THE PRODUCT LABELING (PRODUCT INSERT 0127-1410) PROVIDES WARNING OF POSSIBLE CAP DETACHMENT IN THE PT AND INSTRUCTIONS FOR RETRIEVING.
IT WAS REPORTED BY THE CUSTOMER THAT ON (B)(6) 2010 TWO INCHES OF THE FIBER TIP MELTED AT 143,180 JOULES. NO INJURY WAS REPORTED. AN EXAMINATION, CONDUCTED BY AN AMERICAN MEDICAL SYSTEMS QUALITY ENGINEER, DISCLOSED THAT THE CAP REGION WAS BURNT AND/OR MELTED AND THAT THE CAP REMAINED INTACT AND ATTACHED. THE EXAMINATION ALSO DISCLOSED THAT THE JACKET AND/OR SHRINK TUBE MAY HAVE BEEN MELTED AND/OR BURNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GREENLIGHT ADDSTAT | SURGICAL FIBER | GEX | AMS INNOVATION CENTER-SILICON VALLEY | NA | 944D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |