FDA Adverse Event Malfunction Summary report: N

GREENLIGHT ADDSTAT

MDR report key: 1844131 · Received September 13, 2010

Report

Report Number
2937094-2010-00235
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
March 2, 2010
Report Date
May 5, 2010
Manufacturer
AMS INNOVATION CENTER-SILICON VALLEY
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A FAILURE ANALYSIS REPORT WAS GENERATED BY AN AMERICAN MEDICAL SYSTEMS QUALITY ENGINEER. THE EXAM DISCLOSED THAT THE FIBER CAP HAD WORN OUT. THE CAP REMAINED ATTACHED AND INTACT. THE EXAM ALSO DISCLOSED THAT THE CAP EXHIBITED ALL OR SOME OF CHAR, DEVITRIFICATION, CRATER AND MELTED GLASS AND MINOR BEVEL MELTING. NORMAL WEAR OUT MAY BE ACCELERATED BY TECHNIQUE CAUSING A LACK OF COOLING. THE 2090 FIBER ENDURES A HIGHER POWER, HAS A REFLECTIVE CAP AREA, AND MAY BE SUBJECTED TO MORE HEAT, ESPECIALLY IF COOLING IS BLOCKED BY TISSUE CONTACT. TISSUE CONTACT CREATES CHAR WHICH FURTHER INSULATES THE CAP FROM COOLING AND INCREASES HEAT BY ABSORBING ENERGY. THE HEAT CONTRIBUTES TO DEVITRIFICATION AND ASSOCIATED WEAKENING OF GLASS, MAKING IT SUBJECT TO BREAKAGE FROM MELTING OR MECHANICAL OR THERMAL STRESS. GLUE BURNING INCREASES PRESSURE IN THE CAP. ENERGY REFLECTED BACK AT THE FIBER CAP FROM A SCOPE, SEED, STONE OR OTHERWISE MAY HAVE CONTRIBUTED TO AND/OR CREATED ANY MELTING IN THE CRATER AREA OF THE FIBER CAP. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE USE, ACCELERATED BY TISSUE CONTACT. NO PT INJURY WAS REPORTED. THE PRODUCT LABELING (PRODUCT INSERT 0127-1410) WARNS THAT TISSUE CONTACT OR TISSUE PROBING MAY CAUSE FIBER DAMAGE OR BREAKAGE. THE PRODUCT ALSO WARNS OF POSSIBLE CAP DETACHMENT AND INSTRUCTIONS FOR RETRIEVING.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT ON (B)(6) 2010 THE FIBER'S VAPORIZATION EFFICIENCY SIGNIFICANTLY DIMINISHED AT 18,778 JOULES. NO INJURY WAS REPORTED. AN EXAM, CONDUCTED BY AN AMERICAN MEDICAL SYSTEMS QUALITY ENGINEER, DISCLOSED THAT THE CAP HAD WORN OUT. THE CAP REMAINED ATTACHED AND INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GREENLIGHT ADDSTAT SURGICAL FIBER GEX AMS INNOVATION CENTER-SILICON VALLEY NA 946Z

Patients

Seq Age Sex Outcome Treatment
1 Other