DA VINCI XI
Report
- Report Number
- 2955842-2023-21882
- Event Type
- Malfunction
- Date Received
- January 3, 2024
- Date of Event
- December 12, 2023
- Report Date
- December 12, 2023
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874115343
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- 003
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) DUE TO MULTIPLE GROUNDING ISSUES TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. A RETURN MATERIAL AUTHORIZATION (RMA) HAS BEEN ISSUED REQUESTING TO HAVE THE INTUITIVE SURGICAL, INC. (ISI) DEVICE RETURNED. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.
AN INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) WAS ANALYZED AND FAILURE ANALYSIS INVESTIGATION CONFIRMED AND REPRODUCED THE CUSTOMER REPORTED COMPLAINT. AFTER CONNECTING INSTRUMENTS INTO THE IESU, THE NEUTRAL PORT WAS NOT WORKING, AN ERROR M-0B-31 CAME UP INDICATING A DEFECTIVE INTERNAL COMPONENT.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED HYSTERECTOMY - BENIGN SURGICAL PROCEDURE, CUSTOMER REPORTS, EXPERIENCING A CHECK NEUTRAL ELECTRODE CONNECTION AND M-0B ERROR. THE CUSTOMER STATED THAT THE GROUNDING PAD WAS WORKING AT THE BEGINNING OF THE PROCEDURE, BUT THEN TURNED RED. PRIOR TO CALLING IN THE ISSUE, THE SITE TRIED DIFFERENT GROUNDING PADS AND CLEANING THE AREA WITH NO CHANGE. THE SITE TRIED FOUR GROUNDING PADS. THE SITE ALSO TRIED REBOOTING THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) PRIOR TO CALLING IN THE REPORTED ISSUE. THE TECHNICAL SUPPORT ENGINEER (TSE) INSTRUCTED THE CUSTOMER TO PERFORM A HARD REBOOT ON IESU WITH NO CHANGE. THE SITE REPOSITIONED THE GROUNDING PAD WITH NO CHANGE. THE NURSE WAS ALSO UNABLE TO GROUND THE IESU ON HER ARM AS WELL. THE CUSTOMER STATED THAT SITE DID NOT HAVE ENERGY ACTIVATION CABLE TO CONNECT A FORCE TRIAD GENERATOR. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 559186 | DA VINCI XI | VISION SIDE SYSTEM, RECONDITIONED | NAY | INTUITIVE SURGICAL, INC | 380721-06 | N/A | 00886874115343 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | DA VINCI INSTRUMENTS AND ACCESSORIES. |