FDA Adverse Event
Malfunction
Summary report: N
ACCESS® 2 IMMUNOASSAY SYSTEM
MDR report key: 1844087
·
Received September 24, 2010
Report
- Report Number
- 2122870-2010-00566
- Event Type
- Malfunction
- Date Received
- September 24, 2010
- Date of Event
- August 31, 2010
- Report Date
- September 24, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- MMI
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC PERFORMED PRIOR TO AND AFTER THE EVENT WAS WITHIN THE CUSTOMER'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE VERIFIED HARDWARE PERFORMANCE; ALL TESTING MET SPECIFICATIONS. NO HARDWARE ISSUES WERE NOTED. NO CLEAR ROOT CAUSE COULD BE DETERMINED FOR THIS EVENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING A TROPONIN (ACCUTNI) RESULT, ABOVE THE AMI CUT-OFF, GENERATED BY THE ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT'S SAMPLE. THE RESULT WAS NOT REPORTED OUT OF THE LAB. UPON REPEAT THE RESULTS WERE IN THE NORMAL REFERENCE RANGE. NO REPORTS OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT HAVE BEEN REPORTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | MMI | BECKMAN COULTER INC. | ACCESS® 2 IMMUNOASSAY SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |