BINAXNOW COVID-19 AG SELF TEST 2CT
Report
- Report Number
- 1221359-2024-00020
- Event Type
- Malfunction
- Date Received
- January 3, 2024
- Date of Event
- December 20, 2023
- Report Date
- January 25, 2024
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
D4-UDI: (B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.
D4-UDI: (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 222234 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 222234, TEST BASE PART NUMBER 195-430H / LOT 219670. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 222234 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.
THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL KITTED SWAB. INITIAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A POSITIVE RESULT. ADDITIONAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A NEGATIVE RESULT. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL KITTED SWAB. INITIAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A POSITIVE RESULT. ADDITIONAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A NEGATIVE RESULT. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 587053 | BINAXNOW COVID-19 AG SELF TEST 2CT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 222234 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Male |