FDA Adverse Event Malfunction Summary report: N

HEMOSPRAY ENDOSCOPIC HEMOSTAT

MDR report key: 18440116 · Received January 3, 2024

Report

Report Number
1037905-2024-00007
Event Type
Malfunction
Date Received
January 3, 2024
Date of Event
December 8, 2023
Report Date
April 10, 2024
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
QAU
UDI-DI
00827002246638
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONTINUED: SECTION G: 510K: K200972. INVESTIGATION EVALUATION: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED IN A CLEAR PLASTIC BAG, PROVIDED WITH THE RETURN WAS A LID STOCK FROM THE LOT NUMBER PROVIDED IN THE REPORT. THE LABEL MATCHES THE PRODUCT RETURNED. OUR EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED. ALL COMPONENTS WERE NOT INCLUDED IN THE RETURN (NO CATHETERS WERE RETURNED), THEREFORE A COMPLETE EVALUATION WAS NOT POSSIBLE. THE DEVICE WAS RETURNED WITH THE ON/OFF SWITCH IN THE "ON" POSITION. THE RED ACTIVATION KNOB WAS ENGAGED IN THE HANDLE INDICATING ACTIVATION OF THE CARBON DIOXIDE (CO2) CARTRIDGE. POWDER WAS OBSERVED WITHIN THE DEVICE NOZZLE AND ON THE OUTSIDE OF THE DEVICE. WHEN TESTED AS RETURNED THE DEVICE SPRAYED AS INTENDED. THE CO2 CARTRIDGE AUDIBLY DISCHARGED UPON DEACTIVATION AND THE CARTRIDGE WAS FULLY PUNCTURED. A VISUAL EXAMINATION OF THE O-RING AND LANCE INSIDE THE HANDLE SHOWED BOTH COMPONENTS TO BE POSITIONED CORRECTLY INSIDE THE HANDLE AND THE LANCE TO BE BEVELED. THE INSPECTION OF THE CO2 CARTRIDGE AND REGULATOR (LANCE AND O-RING) CONFIRM THE DEVICE WAS OF THE CURRENT DESIGN. THE FOAM WAS ORIENTED CORRECTLY WITHIN THE HANDLE (SLITS POINTING DOWN TOWARDS THE RED ACTIVATION KNOB). A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: THE LABORATORY EVALUATION OF THE RETURNED DEVICE COULD NOT CONFIRM THE REPORT BECAUSE THE DEVICE SPRAYED AS INTENDED AS RETURNED. THE ROOT CAUSE FOR THE REPORT OF UNABLE TO SPRAY IS UNKNOWN. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. HOWEVER, WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE NO CATHETERS WERE RETURNED FOR EVALUATION. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THIS DEVICE IS WITHIN THE SCOPE OF THE CAPA. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT IS INCLUDED IN THE SCOPE OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

CONTINUED: SECTION G: 510K: K200972. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. A CORRECTIVE ACTION (CAPA) WAS INITIATED TO FURTHER INVESTIGATE DEVICE FAILURE DUE TO BEING UNABLE TO SPRAY POWDER. THIS DEVICE IS WITHIN THE SCOPE OF THE CAPA. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT IS INCLUDED IN THE SCOPE OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING AN ENDOSCOPIC PROCEDURE TO TREAT AN UPPER GI BLEED, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. IT WAS REPORTED THAT AFTER ACTIVATING THE SPRAY HANDLE NOTHING OCCURRED WHEN ATTEMPTING TO FIRE THE SPRAY. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1505972 HEMOSPRAY ENDOSCOPIC HEMOSTAT HEMOSTATIC DEVICE FOR ENDOSCOPIC GASTROINTESTINAL USE QAU WILSON-COOK MEDICAL INC W4583655 00827002246638

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown OLYMPUS ENDOSCOPE, UNKNOWN MODEL.