FDA Adverse Event Other Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1843850 · Received September 21, 2010

Report

Report Number
3004464228-2010-01281
Event Type
Other
Date Received
September 21, 2010
Date of Event
August 22, 2010
Report Date
August 23, 2010
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE POD WAS NOT RETURNED TO THE MFR FOR EVAL. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CONSUMER STATED THAT A KINK WAS SEEN IN THE CANNULA. THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE OMNIPOD USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE EXPERIENCED HIGH BG'S (GREATER THAN 500MG/DL) WITHIN THE FIRST DAY OF ACTIVATING A POD. SHE INDICATED THAT INSULIN WAS LEAKING FROM THE CANNULA AND THAT THE CANNULA WAS KINKED, THOUGH NO ALARM WAS INITIATED. THE POD WILL BE RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30307

Patients

Seq Age Sex Outcome Treatment
1 41 YR Other