FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1843772 · Received September 10, 2010

Report

Report Number
2531779-2010-01237
Event Type
Malfunction
Date Received
September 10, 2010
Date of Event
August 14, 2010
Report Date
August 14, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION FOUND THAT THE PUMP POWERS UP PROPERLY AND THE POWER CIRCUIT DOES NOT DRAW EXCESSIVE CURRENT. OVERHEATING COULD NOT BE DUPLICATED. UPON EXAMINATION, THE PUMP WAS FOUND TO EXHIBIT AN UNREPORTED VISIBLE BATTERY COMPARTMENT CRACK AND CORROSION IN THE BATTERY COMPARTMENT. THE PUMP USER GUIDE INSTRUCTS THE USER TO INSPECT THE PUMP AND CONFIRM THAT IT IS OPERATING PROPERLY. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP.

Description of Event or Problem · 1

THE PATIENT REPORTED THE PUMP WAS HOT TO TOUCH. THE PATIENT DENIED ANY BODILY HARM DUE TO THE TEMPERATURE OF THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 39 YR