FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1843767 · Received September 21, 2010

Report

Report Number
2531779-2010-01304
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
August 23, 2010
Report Date
August 23, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING.

Description of Event or Problem · 1

EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY REVEALED LOSS OF CARTRIDGE DETECTION HAD OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 33 YR