FDA Adverse Event Malfunction Summary report: N

VISERA ELITE II VIDEO SYSTEM CENTER

MDR report key: 18437111 · Received January 3, 2024

Report

Report Number
3002808148-2024-00076
Event Type
Malfunction
Date Received
January 3, 2024
Date of Event
December 11, 2023
Report Date
February 5, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170386022
PMA / PMN Number
K200542
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 2 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, IT IS LIKE THAT THE EVENT REPORTED AS "ERROR E226" WAS CAUSED BY THE BENT PINS IN THE VIDEO SOCKET. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, THE VIDEO PROCESSOR HAS SCOPE COMMUNICATION ERROR E226 . THE ISSUE WAS FOUND DURING PREPARATION FOR USE FOR A DIAGNOSTIC CYSTOSCOPY PROCEDURE. THERE WERE NO REPORTS OF DELAY OR PATIENT HARM.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1550598 VISERA ELITE II VIDEO SYSTEM CENTER VIDEO SYSTEM CENTER FET SHIRAKAWA OLYMPUS CO., LTD. OTV-S200 04953170386022

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown