FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAL PRDGM IN V2.2 CL EN PR

MDR report key: 1843691 · Received September 21, 2010

Report

Report Number
3004209178-2010-82862
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
August 31, 2010
Report Date
August 31, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY. PROBLEM ISOLATED TO THE INTERFACE BOARD.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED AN ERROR ALARM AND HAD A BLANK DISPLAY. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS RESTED AND THE DISPLAY STILL DID NOT RETURN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAL PRDGM IN V2.2 CL EN PR INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR