FDA Adverse Event Malfunction Summary report: N

BD PHOENIX PANEL NMIC-306

MDR report key: 18436329 · Received January 3, 2024

Report

Report Number
1119779-2023-01473
Event Type
Malfunction
Date Received
January 3, 2024
Date of Event
November 25, 2023
Report Date
June 10, 2024
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
00382904492922
PMA / PMN Number
SEE H.10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1. INITIAL REPORTER PHONE #: (B)(6). BD PHOENIX NMIC-306 IS AN ANTIMICROBIAL RESISTANCE PANEL THAT CONSISTS OF A COMBINATION OF THE FOLLOWING 510K NUMBERS: G5. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K163637, K173252, K173523, K181665, K181665, K181665, K181665, K181665, K181665, K181665, K181665, K181665, K190905. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY : THIS COMPLAINT IS FOR LOW MIC MEROPENEM (MEM) WITH PSEUDOMONAS AERUGINOSA WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 3269849. THE CUSTOMER DID NOT PROVIDE ISOLATE RETURNS, PANEL RETURNS BUT PROVIDED BINARY FILES AND PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB RESULTS SHOW LOW MIC MEM RESULTS WITH P. AERUGINOSA WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, THREE RETENTION PANELS OF THE COMPLAINT BATCH WERE INOCULATED WITH IN HOUSE ISOLATE P. AERUGINOSA AND PLACED IN A PHOENIX M50 FOR MEM MIC RESULTS. ALSO, THREE CONTROL PANELS FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH IN HOUSE ISOLATE P. AERUGINOSA AND PLACED IN A PHOENIX M50 FOR MEM MIC RESULTS. DISC DIFFUSION WAS PERFORMED ON THE TESTED ISOLATES TO VERIFY MEM SUSCEPTIBILITY. RESULTS OF THE INVESTIGATION SHOW THE EXPECTED RESULTS FOR MEM. THIS COMPLAINT IS NOT CONFIRMED. THE BREAKPOINTS PUBLISHED IN CLSI M100, 33RD ED /M45-A3 REMAIN UNCHANGED IN THE BD PHOENIX¿, BD EPICENTER¿, AND BD SYNAPSYS¿ SYSTEMS. THE PUD BREAKPOINTS LISTED IN THIS TABLE ARE NOT ALIGNED WITH THE CURRENT CLSI M100 BREAKPOINTS EITHER DUE TO THESE BREAKPOINTS NOT BEING RECOGNIZED BY THE FDA SUSCEPTIBILITY TESTING INTERPRETIVE CRITERIA (STIC) WEBSITE OR PERFORMANCE WITH THE UPDATED BREAKPOINTS HAS NOT YET BEEN EVALUATED INTERNALLY BY BD FOR THE BD PHOENIX ¿ SYSTEM. AS BD PHOENIX¿ CURRENTLY HAS ONLY ONE CLSI INTERPRETATIONAL RULE SET, THIS BREAKPOINT SET IS INFLUENCED BY THE FDA FOR US CUSTOMERS. SEE ATTACHMENT L: BD PHOENIX¿ PUD CLSI MIC BREAKPOINT LISTING FOR A COMPLETE LIST OF THE CLSI MIC BREAKPOINTS PROVIDED IN PUD V7.31. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PHOENIX PANEL NMIC-306 A PATIENT ISOLATE HAD A DISCREPANT RESULT FOR THE DRUG MEROPENEM WHEN COMPARED TO THE ETEST STRIP RESULT. THE PANEL RESULT WAS SENSITIVE WHILE THE ETEST WAS INTERMEDIATE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PHOENIX PANEL NMIC-306 A PATIENT ISOLATE HAD A DISCREPANT RESULT FOR THE DRUG MEROPENEM WHEN COMPARED TO THE ETEST STRIP RESULT. THE PANEL RESULT WAS SENSITIVE WHILE THE ETEST WAS INTERMEDIATE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1549549 BD PHOENIX PANEL NMIC-306 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY LON BECTON DICKINSON & CO. (SPARKS) 3269849 00382904492922

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown