FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522PNAS PRDGM INS SK EN PW

MDR report key: 1843614 · Received September 20, 2010

Report

Report Number
2032227-2010-82697
Event Type
Malfunction
Date Received
September 20, 2010
Date of Event
August 30, 2010
Report Date
August 30, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND DUE TO A CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARMS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED MULTIPLE MOTOR ERROR ALARMS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522PNAS PRDGM INS SK EN PW INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522PNAS

Patients

Seq Age Sex Outcome Treatment
1 61 YR