TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
MDR report key: 1843542
·
Received September 22, 2010
Report
- Report Number
- 2953200-2010-01788
- Event Type
- Injury
- Date Received
- September 22, 2010
- Date of Event
- August 24, 2010
- Report Date
- August 24, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION: RESULTS: ARTERIAL AND VENOUS OCCLUSION, ANGULATED AORTA AND AORTIC NECK. CONCLUSION: ANGULATED AORTA AND AORTIC NECK.
Description of Event or Problem · 1
A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS INSERTED INTO A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK WAS SEVERLY ANGULATED AND 6 CM DISTALLY THE AORTA HAD ANOTHER ANGLE. IT WAS REPORTED THAT AFTER THE BIFURCATED STENT GRAFT WAS IMPLANTED, THE UN-STENTED SEGMENT WAS KINKED AND IT WAS 60% OCCLUDED. AN ANEURX ILIAC EXTENSION LIMB WAS IMPLANTED IN THE UN-STENTED SEGMENT AND RESOLVED THE KINK WITH GOOD FLOW. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00506316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention |