FDA Adverse Event Malfunction Summary report: N

PROXIMATE*ILS CURVED CIR STAPL

MDR report key: 1843538 · Received September 24, 2010

Report

Report Number
3005075853-2010-05487
Event Type
Malfunction
Date Received
September 24, 2010
Report Date
September 1, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4) = WASHER UNCUT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT IF THE INDICATOR IS NOT FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE OR IF THE FIRING HANDLE IS NOT FIRMLY SQUEEZED USING STEADY PRESSURE UNTIL THE FIRING HANDLE IS FULLY PARALLEL TO THE INSTRUMENT HANDLE, STAPLES COULD BE DEPLOYED WITHOUT COMPLETELY FORMING AND WITHOUT CUTTING THE WASHER. FOR MORE INFORMATION, PLEASE REFER TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMA PROCEDURE, INCOMPLETE STAPLE LINE; SUTURED BY HAND. NO FURTHER INFORMATION IS AVAILABLE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS ATRIAL DISLODGED ONE DAY POST-IMPLANT. IN A REVISION PROCEDURE, THE LEAD WAS EXPLANTED AND SUCCESSFULLY REPLACED BY A NEW LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE*ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1