FDA Adverse Event
Injury
Summary report: N
PUMP MMT-722NAB PRDGM INS V2.2 BL EN
MDR report key: 1843439
·
Received September 21, 2010
Report
- Report Number
- 3004209178-2010-82878
- Event Type
- Injury
- Date Received
- September 21, 2010
- Date of Event
- September 4, 2010
- Report Date
- September 7, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED BEING HOSPITALIZED FOR DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 490MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME AND DATE WERE CORRECT. THE BASALS WERE SET INCORRECTLY. THE BOLUS HISTORY REVEALED THAT THE CUSTOMER ENTERED 0.0 UNITS THE DAY OF HER ADMISSION CAUSING HER GLUCOSE LEVEL TO RISE. RAN A FIXED PRIME TEST AND THE INSULIN PUMP PASSED THE TEST. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAB PRDGM INS V2.2 BL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Hospitalization |