FDA Adverse Event Malfunction Summary report: N

VIDEO-OPTIK "ENDOEYE 3D", 30°

MDR report key: 18434222 · Received January 3, 2024

Report

Report Number
9610773-2024-00062
Event Type
Malfunction
Date Received
January 3, 2024
Date of Event
December 19, 2023
Report Date
October 8, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCM
UDI-DI
04042761083492
PMA / PMN Number
K193026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED AND EVALUATED, AND THE CUSTOMER¿S ALLEGATION OF DAMAGE TO THE CHARGED COUPLED DEVICE COVER GLASS WAS NOT CONFIRMED, SHADOWY IMAGE WAS CONFIRMED DUE TO SCRATCHES ON THE CHARGED COUPLED DEVICE COVER GLASS. ADDITIONALLY, IT WAS FOUND THAT THERE WAS A BURST OR BROKEN PLAN GLASS AT THE TUBE. THE INVESTIGATION IS ONGOING, AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. SECTIONS H4 AND H6 WERE UPDATED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS THAT THE VIDEO-OPTIK "ENDOEYE 3D", 30° HAD A SHADOW IN THE IMAGE AND DAMAGE ON THE CHARGED COUPLED DEVICE COVER GLASS. THE EVENT OCCURRED DURING PREPARATION FOR USE IN A THERAPEUTIC PROCEDURE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762637 VIDEO-OPTIK "ENDOEYE 3D", 30° VIDEO SYSTEM CENTER GCM OLYMPUS WINTER & IBE GMBH WA50082A 04042761083492

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown