HD CAMERA HEAD
Report
- Report Number
- 3002808148-2024-30008
- Event Type
- Malfunction
- Date Received
- January 3, 2024
- Report Date
- April 4, 2025
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FET
- UDI-DI
- 04953170336669
- PMA / PMN Number
- K102059
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
E2: NI. G1: MANUFACTURING CONTACT OFFICE FACILITY NAME: SHIRAKAWA OLYMPUS CO., LTD. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
UPDATE B5 THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE SUBJECT DEVICE WAS EVALUATED. DEVICE EVALUATION NOTED THE CUSTOMER REPORTED ISSUE WAS CONFIRMED. ADDITIONAL FINDINGS INCLUDE FLOODING OF THE MAIN BODY, CORROSION OF ELECTRICAL CONTACTS AT CONNECTORS, AND SCOPE MOUNT. MOREOVER, STRIKE MARKS ON ELECTRICAL CONTACTS AT THE CONNECTOR AND WEAR OF MAIN BODY NOTED. THE DEVICE HISTORY RECORDS (DHR¿S) FOR THIS PRODUCT HAVE BEEN REVIEWED. ALL RECORDS INDICATE THAT THE PRODUCT WAS MANUFACTURED AND TESTED IN ACCORDANCE WITH ALL APPLICABLE PROCEDURES AND MET ALL FINAL PRODUCT RELEASE CRITERIA. PER INSTRUCTION FOR USE (IFU), IT STATES, HANDLING AND GENERAL PRECAUTIONS (WARNING) THE ELECTRICAL CONTACTS OF THE VIDEO CONNECTOR AND THEIR SURROUNDINGS SHOULD BE WIPED DRY WITH DRY GAUZE AND CONNECTED TO THE VIDEO SYSTEM CENTER IN A SUFFICIENTLY DRY CONDITION. IN ADDITION, MAKE SURE THAT THE ELECTRICAL CONTACTS ARE FREE OF DIRT BEFORE CONNECTING THEM. USE OF THE EQUIPMENT WITH WET OR DIRTY ELECTRICAL CONTACTS MAY RESULT IN EQUIPMENT FAILURE OR ENDANGER THE SAFETY OF THE PATIENT OR SURGEON. (CAUTION) DO NOT APPLY IMPACT TO THE CAMERA HEAD. DOING SO MAY RESULT IN DAMAGE. DO NOT BUMP OR BEND THE ELECTRICAL CONTACTS OF THE VIDEO CONNECTOR. DOING SO MAY PREVENT CONNECTION TO THE VIDEO SYSTEM CENTER OR CAUSE A POOR CONNECTION." OLYMPUS WILL CONTINUE TO MONITOR COMPLAINTS ABOUT THIS DEVICE.
THIS SUPPLEMENTAL EMDR IS BEING SUBMITTED TO CORRECT THE MANUFACTURING DATE: H4.
CUSTOMER REPORTED THAT THE DEVICE CONNECTOR SECTION WAS CRACKED. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
THIS REPORT IS BEING SUPPLEMENTED BASED ON THE INVESTIGATION COMPLETED BY THE MANUFACTURER.
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 491980 | HD CAMERA HEAD | AUTOCLAVABLE CAMERA HEAD | FET | SHIRAKAWA OLYMPUS CO., LTD. | CH-S190-08-LB | 04953170336669 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |