FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 SYNCHRON® CLINICAL SYSTEM

MDR report key: 1843209 · Received September 24, 2010

Report

Report Number
2050012-2010-00857
Event Type
Malfunction
Date Received
September 24, 2010
Date of Event
August 26, 2010
Report Date
September 23, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE AND REPLACED THE REAGENT SYRINGE VALVES. FSE RAN QC WHICH WAS OKAY AND THE SYSTEM IS OPERATING ACCEPTABLY NOW.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE REAGENT PROBE OF THE UNICEL DXC 800 SYNCHRON CLINICAL SYSTEM HAD A LEAK. NO INJURY WAS REPORTED ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1