INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 1843159 · Received September 23, 2010

Report

Report Number
9611451-2010-00577
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
August 19, 2010
Report Date
August 26, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT DEVICES HAVE NOT BEEN RECEIVED FOR EVALUATION, HOWEVER PHOTOGRAPHS AND A VIDEO WERE PROVIDED BY THE HOSPITAL. RESULTS: THE VIDEO CLEARLY SHOWED THAT ONE OF THE DEVICES WAS LEAKING AT THE SEAL BETWEEN THE SWIVEL ELBOW AND THE SWIVEL WYE. THE PHOTOGRAPHS PROVIDED SHOWED THE BIRD VENTILATOR DISPLAYING A FAILED LEAK TEST. A LOT CHECK REVEALED TWO OTHER COMPLAINTS FOR LOT NUMBER 100310 AND ONE OTHER SIMILAR COMPLAINT FOR LOT NUMBER 100506. CONCLUSION: THE TWO PARTS THAT MAKE UP THE Y-SWIVEL ARE HELD TOGETHER BY A SNAP-FIT, WHICH HAS SOME INHERENT LEAKAGE THAT IS DETECTED AND COMPENSATED FOR BY THE VENTILATOR. IT IS POSSIBLE THAT IF NOT PROPERLY LOCKED INTO PLACE, THE LEAK BETWEEN THE SWIVEL JOINT WAS ENHANCED DURING TRANSPORT OR SETUP DESPITE HAVING PASSED THE LEAK TEST AT THE TIME OF PRODUCTION. ALL CIRCUITS ARE PRESSURE TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCTION LINE. THIS IS AN AUTOMATED PROCESS AND THE CIRCUITS WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT TWO RT235 BREATHING CIRCUITS HAD FAILED THE LEAK TEST ON A BIRD VIP GOLD VENTILATOR IN THE THEIR NEONATAL UNIT. THIS WAS FOUND BEFORE PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT235 100310, 100516

Patients

Seq Age Sex Outcome Treatment
1 BIRD VIP GOLD VENTILATOR