FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 1843110 · Received September 23, 2010

Report

Report Number
2954323-2010-01326
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
September 9, 2010
Report Date
November 7, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

CUSTOMER'S METER WAS RETURNED AND INVESTIGATED. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READING THE CUSTOMER REPORTED WAS FOUND IN METER MEMORY. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING INACCURATE BLOOD GLUCOSE TESTS ON THEIR FREESTYLE LITE METER. THE CUSTOMER REPORTED COMPARING A READING OF 399 MG/DL TO A LAB RESULT OF 170 MG/DL. THE BLOOD TESTS WERE REPORTEDLY COMPLETED WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A THORACOSCOPIC LEFT LOWER LOBECTOMY PROCEDURE, ALTHOUGH ALL THE STAPLES WERE DEPLOYED, THE STAPLES WERE MALFORMED. THE KNIFE MOVED FORWARD PROPERLY. THE EVENT OCCURRED WHEN THE DEVICE WAS FIRED ON THE BRONCHIAL TUBES BETWEEN THE LUNG BASE AND B6. THE TISSUE WAS NOT CALCIFIED, BUT THICK AND HARD. ALTHOUGH THE ARTICULATION LEVER WAS USED, IT WAS NOT FULLY BENT. IT WAS A LITTLE HARD TO GRASP THE CLOSING LEVER, BUT IT COULD BE GRASPED. ALSO, THE FIRING WAS A LITTLE HARD, BUT THE FIRING TRIGGER COULD BE GRASPED. THE EDGE OF THE BRONCHIAL TUBES WAS NOT OPENED JUST AFTER THE FIRING BECAUSE OF PRESSURE BONDING. AFTER THE RESECTED LUNG WAS REMOVED, IT WAS FOUND THAT THE STAPLE LINE HAD OPENED. ALSO, THERE WERE SOME UNFORMED STAPLES INSIDE THE THORACIC CAVITY. THE SMALL INCISION WAS BROADENED OUT A LITTLE AND ADDITIONAL SUTURE WAS PERFORMED WITH A NEEDLE-HOLDER. A NEEDLE COULD BE PRICKED EASILY AT THE TARGET TISSUE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 0935633

Patients

Seq Age Sex Outcome Treatment