FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* BLADELESS TROCAR WITH STABILITY SLEEVE

MDR report key: 1843044 · Received September 23, 2010

Report

Report Number
3005075853-2010-05483
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
August 19, 2010
Report Date
September 2, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DUCKBILL, LEAK TEST FAILURE THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY PROCEDURE, AIR LEAKED AND A HISSING SOUND WAS HEARD. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* BLADELESS TROCAR WITH STABILITY SLEEVE LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1