FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.
MDR report key: 1842966
·
Received September 23, 2010
Report
- Report Number
- 2050012-2010-00831
- Event Type
- Malfunction
- Date Received
- September 23, 2010
- Date of Event
- August 23, 2010
- Report Date
- September 23, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) CLEANED THE GLUCOSE CUP WHICH REDUCED THE IMPRECISION. FSE REPLACED THE GLUCOSE MODULE, WHICH IMPROVED PRECISION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT INCONSISTENT GLUCOSE MODULAR RESULTS WHEN PERFORMING PATIENT SAMPLE POOL OF APPROXIMATELY 100 MG/ML USING UNICEL DXC 800 PRO CHEMISTRY ANALYZER. NO SPECIFIC RESULTS WERE PROVIDED. PATIENT TREATMENT WAS NOT IMPACTED WITH REGARD TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC. | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |