FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.

MDR report key: 1842966 · Received September 23, 2010

Report

Report Number
2050012-2010-00831
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
August 23, 2010
Report Date
September 23, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) CLEANED THE GLUCOSE CUP WHICH REDUCED THE IMPRECISION. FSE REPLACED THE GLUCOSE MODULE, WHICH IMPROVED PRECISION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT INCONSISTENT GLUCOSE MODULAR RESULTS WHEN PERFORMING PATIENT SAMPLE POOL OF APPROXIMATELY 100 MG/ML USING UNICEL DXC 800 PRO CHEMISTRY ANALYZER. NO SPECIFIC RESULTS WERE PROVIDED. PATIENT TREATMENT WAS NOT IMPACTED WITH REGARD TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. CLINICAL CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. N/A N/A