ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2010-01098
- Event Type
- Injury
- Date Received
- September 14, 2010
- Date of Event
- April 16, 2010
- Report Date
- April 29, 2010
- Manufacturer
- ROCHE INSULIN DELIVERY SYSTEMS INC
- Product Code
- LZG
- PMA / PMN Number
- K060876
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- UNKNOWN
Narratives
ON (B)(6) 2010, THE PT'S MOTHER REPORTED THAT ON (B)(6) 2010, THE PT HAD A VIRUS AND WAS VOMITING ALL NIGHT. AT 1 PM ON (B)(6) 2010, THE PT STOPPED VOMITING AND LAID DOWN. THE MOTHER NOTICED THE PT BREATHING HEAVILY AND THE PT WAS UNRESPONSIVE. THE MOTHER TESTED THE PT'S BLOOD GLUCOSE AND IT MEASURED "HI" ON HER BLOOD GLUCOSE MONITOR. SHE INJECTED THE PT WITH 25 UNITS OF INSULIN AND CHANGED ALL OF THE INFUSION DEVICE ACCESSORIES. SHE TESTED THE PT'S BLOOD GLUCOSE AGAIN AND IT STILL MEASURED "HI" AND SHE INJECTED HER WITH ANOTHER 25 UNITS OF INSULIN. SHE SWITCHED THE PT TO THE BACKUP INFUSION DEVICE AND WAS STILL UNABLE TO LOWER HER BLOOD GLUCOSE. AT 11 PM SHE TOOK THE PT TO THE EMERGENCY ROOM. THE HOSPITAL STAFF DISCONNECTED THE PT FROM THE INFUSION DEVICE. HER BLOOD GLUCOSE MEASURED 750 MG/DL AND SHE WAS DIAGNOSED WITH KETOACIDOSIS. SHE WAS TREATED WITH IV FLUIDS AND IV INSULIN. SHE WAS HOSPITALIZED FOR 1 WEEK AND TREATED WITH INSULIN INJECTIONS. WHEN SHE WAS RELEASED FROM THE HOSPITAL SHE RECONNECTED TO THE INFUSION DEVICE PER HER PHYSICIAN AND SHE HAS HAD NO FURTHER ISSUES. THE DAY OF THE INCIDENT THE MOTHER NOTICED INSULIN LEAKING AT THE LUER CONNECTION AND INSULIN LEAKED INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. SHE STATED SHE MAY NOT HAVE PROPERLY TIGHTENED THE INFUSION TUBING. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR EVAL. THE ACCESSORIES IN USE AT THE TIME OF THE INCIDENT HAD BEEN DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK SPIRIT | INSULIN INFUSION PUMP | LZG | ROCHE INSULIN DELIVERY SYSTEMS INC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Hospitalization| O| R | INSULIN| INSULIN INFUSION SET |