FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC

MDR report key: 18427467 · Received January 2, 2024

Report

Report Number
3014704491-2023-00861
Event Type
Malfunction
Date Received
January 2, 2024
Date of Event
October 19, 2023
Report Date
February 29, 2024
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903830831
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW(LOT#3171581): 1) THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN JULY 2023, AND PACKAGED AT R240 PACKAGE LINE IN JULY 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2) REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3) REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLE AND PHOTO HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. 45PSI LEAKAGE TEST IS CARRIED OUT ON THE RETAINED SAMPLE OF THIS BATCH, AND NO LEAKAGE IS FOUND. PLEASE REFER TO ATTACHMENT FOR THE TEST REPORT. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE. AS THE DAMAGE STATE OF THE COMPLAINED SAMPLE CANNOT BE IDENTIFIED, THE ROOT CAUSE OF THE LEAKAGE CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO PAY ATTENTION TO SUCH DEFECTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "PATIENT WAS ADMITTED TO THE HOSPITAL FOR PROSTATE ENLARGEMENT AND WAS GIVEN AN INFUSION TREATMENT ON (B)(6) 2023, WHICH REQUIRED THE USE OF A CLOSED IV INDWELLING NEEDLE, AND WAS FOUND TO BE LEAKING FROM THE STEEL NEEDLE CONNECTION, WHICH WAS REPLACED WITHOUT HARM TO THE PATIENT."

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1227787 BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 3171581 00382903830831

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown