OLYMPUS RESECTION ELECTRODE
Report
- Report Number
- 9610773-2010-00028
- Event Type
- Malfunction
- Date Received
- September 2, 2010
- Date of Event
- August 4, 2010
- Report Date
- August 4, 2010
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- FAS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE REFERENCED IN THIS REPORT WAS DISCARDED BY THE USER FACILITY. THE EXACT CAUSE OF THE PT'S OUTCOME COULD NOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFO BECOMES AVAILABLE LATER A SUPPLEMENTAL REPORT WILL FOLLOW. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. (B)(4). CROSS-REFERENCE MFR REPORT NUMBER 8010047-2010-00170 FOR THE ASSOCIATED DEVICE.
OLYMPUS WAS INFORMED THAT EARLY IN A THERAPEUTIC RESECTION OF PROSTATE (TURP) PROCEDURE THE PHYSICIAN HAD REQUESTED THE IRRIGATION METHOD BE CHANGED. THE PROCEDURE WAS STOPPED, AND WHEN RE-STARTED, THE PHYSICIAN REPORTEDLY ENCOUNTERED A LARGE AMOUNT OF BUBBLES IN THE FIELD OF VIEW THAT MADE VISUALIZATION DIFFICULT. THE PROCEDURE WAS AGAIN STOPPED, AND THE USERS DETERMINED THAT THE NON-OLYMPUS PUMP TUBING HAD BEEN INSTALLED BACKWARDS. THE TUBING WAS CORRECTED AND THE PROCEDURE RESUMED. AFTER APPROX FIVE MINS OF RESECTION, THE PT REPORTEDLY MOVED AND A LOUD POP WAS HEARD. THE PROCEDURE WAS STOPPED AGAIN, AND IT WAS DETERMINED A PERFORATION OF THE BLADDER HAD OCCURRED. THE PT WAS TRANSFERRED TO A DIFFERENT OPERATING ROOM FOR SURGICAL INTERVENTION, AND THE TREATING PHYSICIAN REPORTEDLY DISCOVERED TWO PERFORATIONS IN THE BLADDER. THE PT IS REPORTEDLY DOING FINE FOLLOWING SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS RESECTION ELECTRODE | ELECTRODE | FAS | OLYMPUS WINTER & IBE GMBH | WA22302D | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | CATHETER: MODEL UNK| IRRIGATION PUMP: MODEL UNK| OLYMPUS ELECTROSURIGICAL UNIT: MODEL UES-40 | SN # (B)(6)| OLYMPUS ELECTROSURIGICAL UNIT: MODEL UES-40| SN # (B)(4)| IRRIGATION PUMP: MODEL UNK| CATHETER: MODEL UNK |