FDA Adverse Event Malfunction Summary report: N

BLADE, M4 ROTATABLE RAD 60

MDR report key: 1842655 · Received September 1, 2010

Report

Report Number
1045254-2010-00044
Event Type
Malfunction
Date Received
September 1, 2010
Date of Event
August 3, 2010
Report Date
August 3, 2010
Manufacturer
MEDTRONIC XOMED, INC.
Product Code
EQJ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED FROM THE REPORTER. ANY MISSING OR INCOMPLETE DATA ON THIS FORM 3500A IS THE RESULT OF INFO NOT BEING PROVIDED OR RELEASED BY THE REPORTER, DESPITE MULTIPLE ATTEMPTS TO OBTAIN THE REQUIRED INFO. RECORDS OF THESE ATTEMPTS ARE DOCUMENTED IN THE COMPLAINT FILE. THIS PRODUCT WAS BEING USED FOR TREATMENT, NOT DIAGNOSIS. THE PT IS REPORTED TO BE IN GOOD CONDITION WITH NO ADVERSE SEQUELA AS A RESULT OF THIS EVENT. ALL PORTIONS OF THE BROKEN BUR WERE RETURNED, INDICATING THAT NO DEVICE FRAGMENTS WERE LEFT IN THE PT. VISUAL INSPECTION FOUND NO DAMAGE TO THE TEETH OF THE INNER OR THE OUTER BLADE AND NO DAMAGE TO THE INNER OR OUTER HUBS. VISUAL AND DIMENSIONAL INSPECTION SHOWED THAT THE MIDDLE TUBE FRACTURE AT THE FIRST LOOP OF THE DOVETAIL SPIRAL WRAP CAUSING APPROX 1/2 INCH PORTION OF THE TIP TO BECOME DETACHED. THE "DOVETAIL SPIRAL WRAP" IS THE TOP, CURVED PORTION OF THE MIDDLE TUBE WHERE IT IS DESIGNED TO ALLOW A CURVED TUBE THAT CAN ROTATE. FUNCTIONAL INSPECTION SHOWED THAT THE BLADE WOULD LOAD PROPERLY INTO THE HANDPIECE. IFU STATEMENTS INCLUDE, BUT ARE NOT LIMITED TO: SHOULD A BLADE FRACTURE OCCUR DURING USE, EXTREME CARE MUST BE EXERCISED TO ENSURE THAT ALL FRAGMENTS OF THE BLADE ARE RETRIEVED AND REMOVED FROM THE PT. UNREMOVED BLADE FRAGMENTS MAY CAUSE TISSUE DAMAGE TO THE PT. THIS IS THE FIRST REPORT OF THIS TYPE ON THIS BLADE LOT. WHILE IT IS NOT COMMON FOR A POWERED BUR/BLADE TO BREAK, OUR DATA SHOWS THAT ON RARE OCCASION, A SERIOUS INJURY (SI) OR SURGICAL INTERVENTION HAS OCCURRED DUE TO A BREAK RESULTING IN A DEVICE FRAGMENT. MOST BLADE FRAGMENTS ARE EASILY REMOVED FROM THE PT WITHOUT ADD'L INTERVENTION OR DIRECT HARM AND ALLOW THE SURGICAL PROCEDURE TO PROCEED AS INTENDED. (B)(4).

Description of Event or Problem · 1

DURING SINUS SURGERY, THE BLADE TIP BROKE. THE SURGEON RECOVERED THE PORTION OF THE BLADE FROM THE SINUS WHICH DID NOT RESULT IN A DELAY IN SURGERY. THE BLADE TIP WAS EASILY REMOVED WITHOUT THE USE OF ADD'L EQUIPMENT OR SURGICAL PROCEDURE AND THE ORIGINAL PROCEDURE PROCEEDED AS INTENDED. THERE WAS NO ADVERSE PT SEQUEL REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BLADE, M4 ROTATABLE RAD 60 77EQJ EQJ MEDTRONIC XOMED, INC. 1884016HR 65190400

Patients

Seq Age Sex Outcome Treatment
1 NI