FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 18426478 · Received January 2, 2024

Report

Report Number
9617229-2024-00022
Event Type
Injury
Date Received
January 2, 2024
Date of Event
July 15, 2023
Report Date
March 21, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN/WILL BE REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED AN OPENING ON RADIUS ASSESSED AS FOLD FLAW OPENING. ADDITIONAL OBSERVATIONS: NO OTHER OBSERVATIONS OBSERVED ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED AS THE DEVICE WAS IMPLANTED.

Additional Manufacturer Narrative · 0

VISUAL EVALUATION: DEVICE PHOTOGRAPH(S) FOR THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE-BREAST WAS REQUESTED ON (B)(6) 2024. VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: DEVICE PATCH WITH LOT NUMBER 2112192 AND CATALOG NUMBER TSF365. ¿ RUPTURE- BREAST: OBSERVED, OPENING ON THE DEVICE BUT CANNOT PERFORM AN ASSESSMENT OF THE OPENING AS NO MICROSCOPIC ANALYSIS CAN BE PERFORMED. NO ADDITIONAL OBSERVATIONS ARE PERFORMED. NO FURTHER ACTIONS ARE REQUIRED SINCE NO ISSUE IN THE MANUFACTURING OF THE DEVICE IS OBSERVED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1450561 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2112192

Patients

Seq Age Sex Outcome Treatment
1 51 YR Female Required Intervention