INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2024-00022
- Event Type
- Injury
- Date Received
- January 2, 2024
- Date of Event
- July 15, 2023
- Report Date
- March 21, 2024
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN/WILL BE REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED AN OPENING ON RADIUS ASSESSED AS FOLD FLAW OPENING. ADDITIONAL OBSERVATIONS: NO OTHER OBSERVATIONS OBSERVED ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED AS THE DEVICE WAS IMPLANTED.
VISUAL EVALUATION: DEVICE PHOTOGRAPH(S) FOR THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE-BREAST WAS REQUESTED ON (B)(6) 2024. VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: DEVICE PATCH WITH LOT NUMBER 2112192 AND CATALOG NUMBER TSF365. ¿ RUPTURE- BREAST: OBSERVED, OPENING ON THE DEVICE BUT CANNOT PERFORM AN ASSESSMENT OF THE OPENING AS NO MICROSCOPIC ANALYSIS CAN BE PERFORMED. NO ADDITIONAL OBSERVATIONS ARE PERFORMED. NO FURTHER ACTIONS ARE REQUIRED SINCE NO ISSUE IN THE MANUFACTURING OF THE DEVICE IS OBSERVED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DIAGNOSED BY ULTRASOUND. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1450561 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2112192 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Female | Required Intervention |