FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 18425254 · Received January 2, 2024

Report

Report Number
1024879-2023-00934
Event Type
Malfunction
Date Received
January 2, 2024
Date of Event
December 4, 2023
Report Date
January 10, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903673440
PMA / PMN Number
K030573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY MATERIAL #: 367344. LOT/BATCH#: UNKNOWN . BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 30 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH USED TO DRAW 10 VACUTAINER TUBES, AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.5. PMA / 510(K)#: K220212. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THERE WAS SAMPLE LEAKAGE INTO THE ONE-USE HOLDER. THIS WAS SEEN IN ONE DEVICE. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THERE WAS SAMPLE LEAKAGE INTO THE ONE-USE HOLDER. THIS WAS SEEN IN ONE DEVICE. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1450490 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) UNKNOWN 50382903673440

Patients

Seq Age Sex Outcome Treatment
1 Unknown