FDA Adverse Event
Malfunction
Summary report: N
MICRO SAGITTAL SAW
MDR report key: 1842209
·
Received August 24, 2010
Report
- Report Number
- 1811755-2010-01001
- Event Type
- Malfunction
- Date Received
- August 24, 2010
- Date of Event
- August 3, 2010
- Report Date
- August 3, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE RUNNING ON ITS OWN WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE MOTOR, A WORN BEARING, AND THE SAG HEAD. THOSE PARTS WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE CONTINUED TO RUN ON ITS OWN PRIOR TO THE START OF A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF ANY ADVERSE CONSEQUENCES DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO SAGITTAL SAW | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | HBE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |