FDA Adverse Event
Malfunction
Summary report: N
TERUMO ADVANCED PERFUSION SYSTEM I
MDR report key: 1842167
·
Received August 31, 2010
Report
- Report Number
- 1828100-2010-01237
- Event Type
- Malfunction
- Date Received
- August 31, 2010
- Date of Event
- August 6, 2010
- Report Date
- August 31, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DTQ
- PMA / PMN Number
- K022947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DURING PREPARATION FOR USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED THE O2 SENSOR WAS OUT OF RANGE OF CALIBRATION. THE USER REPORTED THE UNIT WAS POWERED OFF AND ON AND AS A RESULT THE OXYGEN SENSOR CALIBRATED. THE DEVICE WAS USED FOR THE PROCEDURE. THE USER REPORTED THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT AS A RESULT OF THIS RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO ADVANCED PERFUSION SYSTEM I | HEART LUNG CONSOLE | DTQ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 801188 |