FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 1842092 · Received August 26, 2010

Report

Report Number
1811755-2010-01021
Event Type
Malfunction
Date Received
August 26, 2010
Date of Event
August 10, 2010
Report Date
August 10, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE SAG HEAD ASSEMBLY, ROTOR, AND MOTOR, WHICH WERE EACH REPLACED ALONG WITH BEARINGS AND OTHER COMPONENTS. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THE CASE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK