FDA Adverse Event Malfunction Summary report: N

RAD-G

MDR report key: 18420642 · Received December 29, 2023

Report

Report Number
3019388613-2023-00284
Event Type
Malfunction
Date Received
December 29, 2023
Date of Event
January 1, 1901
Report Date
December 6, 2023
Manufacturer
MASIMO - 15750 ALTON PKWY
Product Code
DQA
UDI-DI
00843997000666
PMA / PMN Number
K201770
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACTURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED DEVICE WAS EVALUATED. THE DEVICE POWERED OFF IMMEDIATELY AFTER POWERING ON. THE RAD-G HAD A SHORT INSIDE THE ON/OFF BUTTON WHICH CREATED AN ELECTRICAL SHORT ACROSS THE BUTTON, RESULTING IN THE BUTTON BEING ACTIVATED EVEN WHEN NOT PHYSICALLY PRESSED.

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACURING NARRATIVE: ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

OTHER, OTHER TEXT: MASIMO'S INITIAL REPORT IDENTIFIED THE MODEL AS NUMBER 27977. MODEL NUMBER 9849 IS ASSOCIATED WITH THE GUDID NUMBER. PART NUMBER 27977 IS THE SUBASSEMBLY TO PART NUMBER 9849. THIS SUPPLEMENTAL MDR UPDATES THE MODEL NUMBER TO 9849 AND BRAND NAME TO RAD-G TO ENSURE CORRELATION WITH THE GUDID DATABASE.

Additional Manufacturer Narrative · 0

THIS CORRECTION UPDATES THE UDI NUMBER TO INCLUDE THE DI AND PI FIELDS, ALL NUMBERS, LETTERS, PARENTHESES, AND SYMBOLS.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2245579 RAD-G OXIMETER DQA MASIMO - 15750 ALTON PKWY 9849 00843997000666

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown