RAD-G
Report
- Report Number
- 3019388613-2023-00284
- Event Type
- Malfunction
- Date Received
- December 29, 2023
- Date of Event
- January 1, 1901
- Report Date
- December 6, 2023
- Manufacturer
- MASIMO - 15750 ALTON PKWY
- Product Code
- DQA
- UDI-DI
- 00843997000666
- PMA / PMN Number
- K201770
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL MANUFACTURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED DEVICE WAS EVALUATED. THE DEVICE POWERED OFF IMMEDIATELY AFTER POWERING ON. THE RAD-G HAD A SHORT INSIDE THE ON/OFF BUTTON WHICH CREATED AN ELECTRICAL SHORT ACROSS THE BUTTON, RESULTING IN THE BUTTON BEING ACTIVATED EVEN WHEN NOT PHYSICALLY PRESSED.
ADDITIONAL MANUFACURING NARRATIVE: ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED.
OTHER, OTHER TEXT: MASIMO'S INITIAL REPORT IDENTIFIED THE MODEL AS NUMBER 27977. MODEL NUMBER 9849 IS ASSOCIATED WITH THE GUDID NUMBER. PART NUMBER 27977 IS THE SUBASSEMBLY TO PART NUMBER 9849. THIS SUPPLEMENTAL MDR UPDATES THE MODEL NUMBER TO 9849 AND BRAND NAME TO RAD-G TO ENSURE CORRELATION WITH THE GUDID DATABASE.
THIS CORRECTION UPDATES THE UDI NUMBER TO INCLUDE THE DI AND PI FIELDS, ALL NUMBERS, LETTERS, PARENTHESES, AND SYMBOLS.
THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.
THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.
THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.
THE CUSTOMER REPORTED THE RAD-G "TURNS ON AND THEN TURNS OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2245579 | RAD-G | OXIMETER | DQA | MASIMO - 15750 ALTON PKWY | 9849 | 00843997000666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |