REPAIR CORE IMPACTION DRILL

MDR report key: 1842008 · Received August 26, 2010

Report

Report Number
1811755-2010-01040
Event Type
Malfunction
Date Received
August 26, 2010
Date of Event
July 30, 2010
Report Date
July 30, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH DEBRIS IN THE BEARINGS, WHICH WERE REPLACED. THE ITEM WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN ORAL SURGERY. THERE WAS NO REPORTED PATIENT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR CORE IMPACTION DRILL BONE CUTTING INSTRUMENT AND ACCESSORIES DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK