FDA Adverse Event
Malfunction
Summary report: N
REPAIR CORE IMPACTION DRILL
MDR report key: 1841980
·
Received August 26, 2010
Report
- Report Number
- 1811755-2010-01039
- Event Type
- Malfunction
- Date Received
- August 26, 2010
- Date of Event
- July 30, 2010
- Report Date
- July 30, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH MOTOR CORROSION AND WORN BEARINGS, WHICH WERE REPLACED. THE ITEM WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN ORAL SURGERY. THERE WAS NO REPORTED PATIENT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR CORE IMPACTION DRILL | BONE CUTTING INSTRUMENT AND ACCESORIES | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |