FDA Adverse Event
Malfunction
Summary report: N
TERUMO PERFUSION SYSTEM 8000
MDR report key: 1841849
·
Received September 21, 2010
Report
- Report Number
- 1828100-2010-01498
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- August 29, 2010
- Report Date
- September 21, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- JOR
- PMA / PMN Number
- K915183
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
DURING PREVENTATIVE MAINTENANCE OF THE DEVICE, THE FIELD SERVICE TECH REPORTED, THE SHOULDER SCREW OF THE ROLLER PUMP WAS UNABLE TO BE TORQUED TO 17 INCHES DUE TO THE SLIDING MECHANISM BEING STRIPPED. NO OTHER DETAILS REGARDING THE NATURE OF THE EVENT WERE PROVIDED. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE OF THE DEVICE, THERE WAS NO PT INVOLVEMENT DURING THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO PERFUSION SYSTEM 8000 | ROLLER PUMP | JOR | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 16402 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |