I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2023-00177
- Event Type
- Malfunction
- Date Received
- December 29, 2023
- Date of Event
- December 20, 2023
- Report Date
- January 23, 2024
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749001917
- PMA / PMN Number
- K053110NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 17-JAN-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AL, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H23247.
ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT POTASSIUM RESULT OF >9.0 MMOL/L ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD, RESULT, SAMPLE: I-STAT >9.0, MMOL/L, A; I-STAT, 4.2 MMOL/L, A; LAB, 4.2, MMOL/L, NI. COLLECTION/TEST TIMES, LAB METHOD WERE NOT PROVIDED AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1842133 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE, | JGS | ABBOTT POINT OF CARE | NA | H23247 | 10054749001917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |