FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 18418054 · Received December 29, 2023

Report

Report Number
2245578-2023-00177
Event Type
Malfunction
Date Received
December 29, 2023
Date of Event
December 20, 2023
Report Date
January 23, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749001917
PMA / PMN Number
K053110NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 17-JAN-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AL, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H23247.

Description of Event or Problem · 0

ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT POTASSIUM RESULT OF >9.0 MMOL/L ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD, RESULT, SAMPLE: I-STAT >9.0, MMOL/L, A; I-STAT, 4.2 MMOL/L, A; LAB, 4.2, MMOL/L, NI. COLLECTION/TEST TIMES, LAB METHOD WERE NOT PROVIDED AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Description of Event or Problem · 0

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1842133 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE, JGS ABBOTT POINT OF CARE NA H23247 10054749001917

Patients

Seq Age Sex Outcome Treatment
1 Unknown