MATRIX DETACHABLE COILS
Report
- Report Number
- 2939204-2010-00958
- Event Type
- Injury
- Date Received
- September 22, 2010
- Date of Event
- December 8, 2009
- Report Date
- September 9, 2010
- Manufacturer
- BOSTON SCIENTIFIC - CORK
- Product Code
- HCG
- PMA / PMN Number
- UNKNOWN
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
PMA# OR 510K# - K050700.
PMA# OR 510K# - K050700. CONCLUSION: ANTICIPATED PROCEDURAL COMPLICATION. THE DEVICE REMAINS IMPLANTED SO WAS NOT AVAILABLE FOR EVALUATION. FROM THE INFORMATION PROVIDED THERE IS NO INDICATION THAT THERE WAS ANY DEVICE MALFUNCTION, NONCONFORMANCE OR MISUSE THAT CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE LABELING FOUND THAT THE DIRECTIONS FOR USE NOTES THAT MULTIPLE PROCEDURES MAY BE REQUIRED TO OCCLUDE AN ANEURYSM. THEREFORE IT WAS DETERMINED THAT THE REPORTED EVENT WAS AN ANTICIPATED PROCEDURAL COMPLICATION.
AT 460 DAYS POST INDEX PROCEDURE, THE PATIENT UNDERWENT A SECOND PROCEDURE TO OCCLUDE THE ANEURYSM. A STENT AND TWENTY THREE COILS WERE SUCCESSFULLY IMPLANTED. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.
AT 460 DAYS POST INDEX PROCEDURE, THE PATIENT UNDERWENT A SECOND PROCEDURE TO OCCLUDE THE ANEURYSM. A STENT AND TWENTY THREE COILS WERE SUCCESSFULLY IMPLANTED. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MATRIX DETACHABLE COILS | DEVICE, NEUROVASCULAR EMBOLIZATION | HCG | BOSTON SCIENTIFIC - CORK | UNK536 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Required Intervention | TWO (B)(4) - MATRIX2 360 SR COILS| (B)(4) - MATRIX2 ULTRASOFT SR COIL| THREE (B)(4) - MATRIX2 360 SR COILS| (B)(4) - MATRIX2 ULTRASOFT SR COIL| (B)(4) - MATRIX2 360 SR COIL| (B)(4) - MATRIX2 360 SR COIL |