FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1841642 · Received September 16, 2010

Report

Report Number
2953200-2010-01761
Event Type
Injury
Date Received
September 16, 2010
Date of Event
August 17, 2010
Report Date
August 17, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: (ENDOLEAK), (ANGULATED AORTIC NECK AND TORTUOUS ILIAC ARTERIES). CONCLUSIONS: (ANGULATED AORTIC NECK AND TORTUOUS ILIAC ARTERIES).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 5.9 CM ABDOMINAL AORTIC ANEURYSM APPROXIMATELY ONE MONTH AGO. THE ILIAC VESSELS WERE TORTUOUS AND CALCIFIED. THE AORTIC NECK IS GREATER THAN 30 MM IN LENGTH, MODERATELY ANGULATED, HAD MILD CALCIFICATION, AND 25 MM IN DIAMETER PROXIMALLY AND 27 MM IN DIAMETER DISTALLY. IT WAS REPORTED THAT AFTER BALLOONING THE STENT GRAFTS, A POST-ANGIOGRAM DEMONSTRATED A PROXIMAL TYPE I ENDOLEAK (MFR REPORT # 2953200-2010-01761). IT APPEARED THAT THE NECK ANGULATION WAS PREVENTING SEAL ON THE RIGHT SIDE, AS THE GRAFT SAT 1 CM BELOW THE RIGHT RENAL ARTERY; CONTRAST COULD BE SEEN BEYOND THE GRAFT AT THAT LOCATION. AS APPROXIMATELY 3-5 MM WERE AVAILABLE BELOW THE LEFT RENAL, THE PHYSICIAN DECIDED TO PLACE AN AORTIC CUFF FROM THE LEFT SIDE TO BETTER MATCH THE ANGLE OF THE AORTA. HOWEVER, ACCURATE PLACEMENT WAS DIFFICULT DUE TO THE TORTUOUS ILIAC ARTERIES, AND CONSEQUENTLY, THE LEFT RENAL ARTERY WAS ENCROACHED UPON BY THE AORTIC CUFF (MFR REPORT # 2953200-2010-01762). ATTEMPTS TO STENT THE LEFT RENAL ARTERY WERE UNSUCCESSFUL, AND THE PHYSICIAN ELECTED NOT TO INTERVENE ANY FURTHER. THE LEFT RENAL ARTERY WAS FILLING ON THE ANGIOGRAM, AND THE PROXIMAL TYPE I ENDOLEAK HAD RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00500811

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention