TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-01759
- Event Type
- Injury
- Date Received
- September 16, 2010
- Date of Event
- August 17, 2010
- Report Date
- August 17, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (GRAFT MIGRATION, ENDOLEAK), (CONICAL AORTIC NECK WITH ANGULATION). CONCLUSION: (CONICAL AORTIC NECK WITH ANGULATION).
A TALENT STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 2 YEARS AGO. ANEURYSM MORPHOLOGY AT THE TIME OF IMPLANT WAS NOT REPORTED. THE AORTIC NECK WAS CONICAL IN SHAPE, RANGING FROM 27 MM IN DIAMETER AT THE RENAL ARTERIES AND 36 MM IN DIAMETER AT 15 MM BELOW. THE NECK WAS ALSO ANGULATED 45 DEGREES LATERALLY AND THEN TOOK A 45 DEGREE TURN IN THE OTHER DIRECTION. IT WAS REPORTED THAT THE BIFURCATED STENT GRAFT HAD MIGRATED 4 MM DISTALLY AND HAD FALLEN INTO THE ANEURYSM SAC APPROXIMATELY ONE MONTH AGO, RESULTING IN A PROXIMAL TYPE I ENDOLEAK. THE MIGRATION WAS ATTRIBUTED TO THE NECK ANGULATION. THE PHYSICIAN ELECTED TO INTERVENE BY PLACING TWO TALENT THORACIC 36 MM AORTIC CUFFS TO BUILD THE DEVICE UP FROM THE FLOW DIVIDER TO THE RENAL ARTERIES, WHICH SUCCESSFULLY RESOLVED THE ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR | NA | V00183209 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |