FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 5ML LS

MDR report key: 18415597 · Received December 28, 2023

Report

Report Number
2243072-2023-02253
Event Type
Malfunction
Date Received
December 28, 2023
Date of Event
December 1, 2023
Report Date
January 19, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CURRENT CONTROL THERE IS VISUAL INSPECTION FOR EXCESSIVE / NO SILICONE AND QA OUTGOING INSPECTION THERE IS PLUNGER BREAKOUT AND SUSTAINING FORCE TO CHECK FOR TIGHTNESS. DUE TO NO SAMPLE, NO PHOTO WAS RETURNED, ROOT CAUSE COULD NOT BE DETERMINED. COMPLAINT WILL BE MONITORED AND REOPENED IF SAMPLE IS RETURNED.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

ONE OF OUR ANESTHETIST ADVISED TODAY THAT DURING AN EYE INJECTION WITH A 5ML SYRINGE FROM BD WAS VERY STIFF THE PLUNGER WAS AND HARD TO PULL AND PUSH. HE THEN CHANGED SYRINGE TO ANOTHER BRAND AND FOUND THAT IT WAS THE SYRINGE AND NOT THE EYE PRESSURE. WE THEN PROCEED TO OPEN ANOTHER BD SYRINGE TO DIS COVER THAT THE Y WERE VERY STIFF AND HARD TO PULL AND PUSH THE PLUNGER AND COMPARE D TO ANOTHER BRAND. REF: 302130. BATCH LOT: 2141756. MANUFACTURE DATE : 2022-06. EXPIRY DATE: 2027-05.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE 5ML LS PLUNGER MOVEMENT WAS DIFFICULT THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ONE OF OUR ANESTHETIST ADVISED TODAY THAT DURING AN EYE INJECTION WITH A 5ML SYRINGE FROM BD WAS VERY STIFF THE PLUNGER WAS AND HARD TO PULL AND PUSH. HE THEN CHANGED SYRINGE TO ANOTHER BRAND AND FOUND THAT IT WAS THE SYRINGE AND NOT THE EYE PRESSURE. WE THEN PROCEED TO OPEN ANOTHER BD SYRINGE TO DIS COVER THAT THE Y WERE VERY STIFF AND HARD TO PULL AND PUSH THE PLUNGER AND COMPARE D TO ANOTHER BRAND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1400403 BD SYRINGE 5ML LS SYRINGE FMF BECTON DICKINSON 2141756

Patients

Seq Age Sex Outcome Treatment
1 Unknown